This is the metadata for: MUMBA_NOx_CSIRO_2013-01-31_2013-02-16 (doi:10.1594/PANGAEA.871994) Measurement: NO and NO2 Instrument: Thermo TEI42i chemiluminescence instrument with blue light converter Owner: CSIRO Campaign: MUMBA Location: Main measurement site - Nissan Hut Kids Campus / Campus East, Wollongong (-34.3972, 150.8996) Dates of operation: 31-Jan-2013 to 15-Feb-2013 Original measurement time resolution: 1 minute Averaging time: 1 hour Units: ppb Inlet/Installation detail (inc. flow rate if relevant): Ambient air sampled with 3/8" PTFE bypass flow, 1/4" PTFE inlet, not heated, inline Teflon filter, max. residence time 3 seconds, shared with NOx (UOW), O3 and SIFT systems which all take their samples downstream of the instrument Data/sample collection detail: blue light converter - NO2 measurements are free from interferences from nitrates Data/sample analysis detail (inc. references if needed): NO2 mixing ratios calculated as the difference between NOx and NO mixing ratios (NO2 = NOx - NO) Calibration: NO only - February 2013 Accuracy: NO +- 1% Issues: none reported Corrections: No corrections were applied - see the data file for the UOW NOx instrument for the magnitude of the correction for the inlet line reaction (NO + O3-> NO2 + O2), calculated assuming the entire sampling line was dark Uncertainty: timeshift +-20s Missing data reported as empty fields File headers: date yyyy-mm-ddTHH:MM:SSz NO (ppb) mean hourly NO in mixing ratio by volume NO2 (ppb) mean hourly NO2 in mixing ratio by volume Contact(s): Elise-Andree Guerette eag873@uowmail.edu.au Ian Galbally ian.galbally@csiro.au Dagmar Kubistin dagmar.kubistin@dwd.de