This is the metadata for: MUMBA_NOx_UOW_2012-11-21_2013-02-15 (doi:10.1594/PANGAEA.871995) Measurement: NO and NO2 Instrument: Thermo TEI42i chemiluminescence instrument Owner: University of Wollongong Campaign: MUMBA Location: Main measurement site - Nissan Hut Kids Campus / Campus East, Wollongong (-34.3972, 150.8996) Dates of operation: 21-Dec-2012 to 15-Feb-2013 Original measurement time resolution: 1 minute Averaging time: 1 hour Units: ppb Inlet/Installation detail (inc. flow rate if relevant): Ambient air sampled with 3/8" PTFE bypass flow, 1/4" PTFE inlet, not heated, inline Teflon filter, max. residence time 3 seconds, shared with O3 and SIFT systems which take their samples downstream of NOx instrument; from 31/01/2013 CSIRO NOx box connected upstream Data/sample collection detail: Molybdenum converter; NO2 measurements are influenced by interferences caused by nitrates with an unkown conversion efficiency Data/sample analysis detail (inc. references if needed): NO2 mixing ratios calculated as the difference between NOx and NO mixing ratios (NO2 = NOx - NO) Calibration: NO only - February 2013 Precision (60 seconds): NO 0.027 ppb, NO2 0.045 ppb, adapted from NO Accuracy: NO +- 1% Issues: none reported Corrections: a correction for the inlet line reaction (NO + O3-> NO2 + O2) was applied to the original 60 second data, assuming the entire sampling line was dark (median: 0.007 ppb, mean: 0.015 ppb) - mean hourly values are reported Uncertainty: uncertainty due to the difference in calibrations prior and post to MUMBA campaign for NO: +-13%, timeshift +-20s Detection Limit (2-sigma, 60 seconds): NO 51 pptv, NO2 91 pptv, NOx 91 pptv Data below detection limit reported as: included Missing data reported as: NA File headers: date yyyy-mm-ddTHH:MM:SSz NO (ppb) mean hourly NO in mixing ratio by volume precision NO (ppb) mean hourly precision for NO in mixing ratio by volume NO2 (ppb) mean hourly NO2 in mixing ratio by volume precision NO2 (ppb) mean hourly precision for NO2 in mixing ratio by volume inline reaction (ppb) mean hourly correction applied for inlet line reaction (NO + O3-> NO2 + O2) in mixing ratio by volume data version: 18-Sep-2014 Contact(s): Dagmar Kubistin dagmar.kubistin@dwd.de