This is the metadata for: MUMBA_O3_2012-12-21_2013-02-15 (doi:0.1594/PANGAEA.871996) Measurement: Ozone Instrument: Thermo TEI49i UV absorption instrument Owner: UOW Campaign: MUMBA Location: Main measurement site - Nissan Hut Kids Campus / Campus East, Wollongong (-34.3972, 150.8996) Dates of operation: 21-Dec-2012 to 15-Feb-2013 Original measurement time resolution: 1 minute Averaging time: 1 hour Units: ppb Inlet/Installation detail (inc. flow rate if relevant): Ambient air sampled with 3/8" PTFE bypass, 1/4" PTFE inlet, not heated, inline Teflon filter, max. residence time 3 seconds. Inlet shared with NOx and SIFT system which took their sample upstream and downstream of O3 instrument, respectively Data/sample collection detail: Data/sample analysis detail (inc. references if needed): Calibration: May 2013 Estimated inlet loss: 1% Precision: +- 1 ppb Accuracy: +- 2.1% - mainly determined by the uncertainty of the O3 absorption cross section (Viallon et al 2012) Issues: none reported Corrections (inc. blank corrections - this may have been covered in analysis detail): Uncertainty: timestamp - less then +- 20 s, drift - uncertainty due to the difference in calibrations prior and post to MUMBA campaign (see file) Detection Limit: 0.5 ppb (2-sigma, 60 seconds measurement) Missing or invalid data reported as empty fields File headers: date yyyy-mm-ddTHH:MM:SSz median_O3 (ppb) median of hourly O3 in volume mixing ratio mean_O3 (ppb) mean of hourly O3 in volume mixing ratio std_O3 (ppb) standard deviation of hourly mean O3 in volume mixing ratio uncertainty_drift_median_O3 (ppb) uncertainty on median caused by difference in prior and post calibration uncertainty_drift_mean_O3 (ppb) uncertainty on mean caused by difference in prior and post calibration uncertainty_drift_median_O3 (ppb) uncertainty on standard deviation caused by difference in prior and post calibration no_points number of points used to calculate median and mean O3 values data version: 14-Aug-2013 Contact(s): Dagmar Kubistin dagmar.kubistin@dwd.de References Viallon, J., P. Moussay, F. Idrees, R. Wielgosz and C. Zellweger (2012). "Final report, ongoing key comparison BIPM.QM-K1: Ozone at ambient level, comparison with EMPA (June 2012)." Metrologia 49(1A): 08011, doi:10.1088/0026-1394/49/1A/08011