This is the metadata for: MUMBA_SMPS_2013-01-16_2013-02-18 (doi:10.1594/PANGAEA.872000) Measurement: Particle number size distribution (~14 nm to ~660 nm) Instrument: TSI Scanning Mobility Particle Sizer (SMPS, Model 3080) with long Differential Mobility Analyser (DMA, TSI 3071A) and CPC (TSI 3010) (64 channels) Owner: CSIRO Campaign: MUMBA Location: Main measurement site - Nissan Hut Kids Campus / Campus East, Wollongong (-34.3972, 150.8996) Dates of operation: 16-Jan-2013 to 15-Feb-2013 Original measurement time resolution: samples for ~2.5 minutes every 5 minutes Averaging time: 1 hour Units: diameter: nm particle concentrations: dN/dLogDp particles cm-3 Inlet/Installation detail: Air was sampled through an 8 m copper bypass running at a flow rate of ~15L/minute. Inlet height was ~10m. This bypass was shared between the CPC, CCNC and SMPS instruments. Data/sample collection detail: aerosol flow was 0.30 +- 0.03 L/minute and sheath flow was 3.0 +- 0.3 L/minute Calibration: Sizing accuracy of the SMPS system was checked using PolyStyrene Latex (PSL) solution with errors being less than 2 % Issues: The instrument was shut down due to excessive heat on 18-20-Jan-2013 Corrections: Original timestamp of data collected from 16-Jan-2013 until 18-Jan-2013 was corrected from GMT+11 to GMT+10 Data was corrected for inlet flow rates (20%) and inlet losses (0.9 transmission factor) and filtered for logged poor periods. Uncertainty: +-10% (Wiedensohler et al 2012) Missing data reported as empty fields File headers: date yyyy-mm-ddTHH:MM:SSz ... the next 107 columns have headers that are the midpoint particle diameter (in nm) for the corresponding bin, the particle numbers are normalised for the number of channels (64) ... TotalConc - total number of particles Geometrian Median Diameter - median diameter of the particle distribution Contact(s): Ruhi Humphries ruhi.humphries@csiro.au Melita Keywood melita.keywood@csiro.au References Wiedensohler, A., W. Birmili, A. Nowak, A. Sonntag, K. Weinhold, M. Merkel, B. Wehner, T. Tuch, S. Pfeifer, M. Fiebig, A. M. Fjäraa, E. Asmi, K. Sellegri, R. Depuy, H. Venzac, P. Villani, P. Laj, P. Aalto, J. A. Ogren, E. Swietlicki, P. Williams, P. Roldin, P. Quincey, C. Hüglin, R. Fierz-Schmidhauser, M. Gysel, E. Weingartner, F. Riccobono, S. Santos, C. Grüning, K. Faloon, D. Beddows, R. Harrison, C. Monahan, S. G. Jennings, C. D. O'Dowd, A. Marinoni, H. G. Horn, L. Keck, J. Jiang, J. Scheckman, P. H. McMurry, Z. Deng, C. S. Zhao, M. Moerman, B. Henzing, G. de Leeuw, G. Löschau and S. Bastian (2012). "Mobility particle size spectrometers: harmonization of technical standards and data structure to facilitate high quality long-term observations of atmospheric particle number size distributions." Atmos. Meas. Tech. 5(3): 657-685. doi:10.5194/amt-5-657-2012